9-4
2790 SourceMeter
®
Switch System Reference Manual
Front panel scanning (7702 only)
When a simple scan using the 7702 module is configured, the present limit values and
state will apply to all channels in the scan. When an advanced scan is configured, each
channel can have its own unique limits configuration. Details to configure and run a scan
are provided in
.
Remote scanning
For remote programming, the <clist> parameter is used to configure channels for a 7702
scan (
). The <clist> parameter can also include memory pattern locations to
allow scanning through 7751/7752/7753 channels (see
).
Basic limits operation
The limits configuration is the same for all functions. For example, if a reading limit is set
to 1, that will equate to 1V for a voltage function, 1A for a current function, and 1
Ω
for an
ohms function.
Setting limits
1.
Press SHIFT and then LIMITS to display the high limit for Limit 1 (HI1).
2.
Use
,
,
, and
to key in the HI1 limit and press ENTER.
3.
When editing a reading, use the range designator (
, K, or M) as a multiplier.
With the cursor on the range designator, each press of
or
will increase or
decrease the reading by a factor of 10.
4.
Key in the low limit for Limit 1 (LO1) and press ENTER.
5.
Key in the high limit for Limit 2 (HI2) and press ENTER.
6.
Key in the low limit for Limit 2 (LO2) and press ENTER. The instrument will
return to the normal measurement state.
Beeper settings
The beeper is configured from the OUTPUT menu (shown in
) as follows:
1.
Press SHIFT and then OUTPUT.
2.
Use the
or
key to display the present beeper (BEEP) setting: NEVER,
INSIDE, or OUTSIDE.
3.
Press
to position the cursor on the present beeper setting, use the
or
key
to display the desired setting and press ENTER. The instrument will return to the
normal measurement state.
klqb
Remote programming cannot be used to set the beeper. It can only be set from
the front panel.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...