7-12
Model 2790 SourceMeter
®
Switch System Reference Manual
Figure 7-3
Scan configuration flowchart
klqb
The instrument is always configured to run a scan. On power-up, each available
channel uses the power-on default setup. For example, for factory power-on
default settings and a Model 7702 installed in slot 2, the instrument will scan
channels 201 through 240 when the scan is run. See
for details on
power-on default settings.
There are two scan configurations: simple and advanced. When you configure the simple
scan, the instrument uses the present instrument setup for each channel in the scan. For the
advanced scan, each channel can have its own unique setup. As explained in
, there is a user-set delay (auto or manual) that is in effect for both the
simple and advanced scan.
Channel setup considerations
Rel
— In order to use an acquired rel value for an advanced scan channel, the rel value has
to be acquired with the instrument in the normal measurement state. Details to set rel for
scan channels are provided in
. Scanning examples (front panel and
remote programming) at the end of this section demonstrate how to set rel values for scan
channels.
Filter
— The moving filter cannot be used in a scan; only the repeat filter can be used. If
you configure a channel (or channels) to use the moving filter, the filter will be off when
the scan is run. See
for details on filter.
SHIFT
CONFIG
Simple
Advanced
Reset
Min Chan
Max Chan
Setup
Imm Scan?
Timer?
Rdg Cnt
Timer?
Rdg Cnt
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...