3-34
Model 2790 SourceMeter
®
Swich System Reference Manual
Commands
1
Description
Default Ref
CONT function
[SENSe[1]]
Optional root command.
:CONTinuity:THReshold <NRf>
Set continuity threshold in ohms;
<NRf> = 1 to 1000.
10
SYSTem:BEEPer <b>
Enable/disable beeper; <b> = ON or OFF.
ON
Set temperature measurement units
UNIT:TEMPerature <name>
Set temperature units; <name> = C, CEL,
F, FAR, or K.
f*
Trigger and retrieve readings
g*
INITiate:CONTinuous <b>
Enable/disable continuous initiation;
<b> = ON or OFF.
INITiate[:IMMediate]
Trigger one or more measurements.
g*
[SENSe[1]]
Optional root command.
:DATA[:LATest]?
Returns the last reading string.
g*
:DATA:FRESh?
Returns the last “fresh” reading string.
g*
FETCh?
Return reading(s).
g*
READ?
Trigger and return reading(s).
g*
Channel list parameter (7702):
<clist> = (@SCH)
where:
S = Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Channel list parameter (7751/7752/7753, DCV,
Ω
2,
Ω
4 only):
<clist> = (@Mn)
where:
M = Memory pattern command
n = Location number (1 to 40)
*Refer to the paragraphs following this table for more information.
1
The <clist> parameter is used for the module to configure one or more channels for a scan.
Table 3-5 (cont.)
Basic measurement commands
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...