15-8
Model 2790 SourceMeter
®
Switch System Reference Manual
:ROUTe
ROUTe subsystem (continued):
:SCAN
Path to configure scan (7702 only):
✓
[:INTernal] <clist>
Specify list of channels to be scanned.
✓
[:INTernal]?
Query scan list.
✓
:TSOurce <list>
Select trigger source to start the scan
(IMM, or HLIM1, HLIM2, LLIM1,
and LLIM2).
IMM
:TSOurce?
Query trigger source for scan.
:NVOLatile <b>
Enable or disable nonvolatile memory
for scanning (autoscan).
(Note 2)
:NVOLatile?
Query nonvolatile memory setting.
:LSELect <name>
Enable (INTernal) or disable (NONE)
scan.
NONE
:LSELect?
Query state of scan.
:MEMory
Path to memory pattern commands:
(Note 4)
[:CHANnels] <n>,<clist>
Define channel pattern for memory
location <n> (1-40).
(Note 5)
[:CHANnels]?<n>
Query channel pattern for memory <n>.
:SOURce
:LEVel <n>,<NRf>, <clist>
Assign source values to channels.
(Note 5)
:LEVEL? <n>
Query source values.
:DELay <n>, <NRf>
Assign a variable delay to memory
<n>.
(Note 5)
:DELay? <n>
Query delay from memory <n>.
:RECall <n>
Execute memory pattern <n>.
:READ
[:STATe] <n>, <b>
Enable/disable reading for memory
<n>.
(Note 5)
:CLEar <n>
Clear out memory pattern <n>.
:ALL
Clear out all memory pattern locations.
Table 15-4 (cont.)
ROUTe command summary
Command
Description
Default
parameter
Ref
SCPI
Notes:
1. Default value depends on which switching module is installed.
2. Not affected by *RST and SYSTem:PRESet. Front panel factory default is OFF.
3. Not affected by *RST and SYSTem:PRESet. Interval set to 15 minutes at the factory.
4. Memory patterns available in firmware revision A04 and higher. 40 memory locations available.
5. Memory pattern locations not affected by *RST or SYSTtem:PRESet. Use ROUT:MEM:CLE to clear memory.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...