Recommended test equipment
summarizes the test equipment required for the manufacturing calibration
steps. Note that the equipment depends on the firmware revision level. See
equipment needed to complete the comprehensive calibration steps.
Extender board preparation
Before performing manufacturing calibration, short the output HI, LO, SHI, and SLO
terminals of the Keithley 7797 Calibration/Extender/Test Board together using clean,
solid copper wires. These connections will form a low-thermal short necessary for the
manufacturing calibration procedure. The Calibration Extender Test Board should then
be installed in scanner Slot #1.
Test equipment connections
Connections for steps requiring the calibrator are shown in
shows function generator connections. Refer to the test procedures for specific steps.
Unlocking manufacturing calibration
To unlock manufacturing calibration, press and hold in the OPEN key while turning on the
power.
Table 17-13
Recommended equipment for manufacturing calibration
Firmware revision A03 and lower
Firmware revision A04 and higher
Stanford Research Systems
DS345 Function Generator:
1V RMS, 3Hz, ±5ppm
1V RMS, 1kHz, ±5ppm
Stanford Research Systems
DS345 Function Generator:
1V RMS, 1kHz, ±5ppm
Keithley Model 2001 Digital Multimeter:
1V, 3Hz AC, ±0.13%
Fluke 5700A Calibrator:
1V RMS, 10Hz, ±260ppm
Keithley 7797 Calibration/Extender Board
Keithley 7797 Calibration/Extender Board
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...