6 Panoramic and 3D calibrations and adjustments
62
performed as a support for the installer. All calibrations
must be done until the system informs they have passed.
The exclamation marks disappear when a passed
calibration has been saved.
NOTICE! After calibrations place all the needed patient
positioning accessories back to the chin rest.
NOTICE! Press Save on GUI after each calibration.
6.2 When to calibrate the unit
The unit must be calibrated and, if necessary, adjusted at
regular intervals in accordance with national regulations
regarding the use, maintenance and service of dental x-ray
devices that are in force in the country in which the unit is
installed.
In any event, the unit must be completely re-calibrated at
least once a year.
The unit must also be fully or partially calibrated when
parts are replaced. The calibration steps for the following
replaced parts are as follows:
Tubehead
After replacing the tubehead the unit must be completely
recalibrated.
NOTICE! Also perform Quality Check when you replace
the tubehead.
R3800 X-ray generator
After replacing R3800 X-ray generator preheat and mA
calibration must be redone.
Collimator assembly
After replacing or adjusting the collimator assembly redo all
applicable collimator, pixel and geometry calibrations as
well as the quality checks.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...