7 Cephalometric calibration and alignment
97
7.2.4
Mechanical adjustment program
Mechanical adjustments of primary and secondary
collimator, as well as adjustment of the cephalostat fixation
angle, are guided by a mechanical adjustment program.
This program gives an image with three exposed areas of
interest (marked with arrows in the figure).
•
The leftmost area is exposed with the cephalomet-
ric collimator having reduced height. This area is
used for adjusting the vertical position of the X-ray
beam (primary collimation vertical position).
•
The area in the middle is exposed with the full
height cephalometric collimator. This area is used
for checking X-ray beam integrity, and shall be
equally exposed in the whole height of the image.
•
The rightmost area is the beam after secondary
collimation. This area is used for checking second-
ary collimator vertical position and tilt. This area
being too far left or right in the image indicates that
the fixation angle of the cephalostat is incorrect
and that the cephalostat must be turned.
Example of a Cephalostat Mechanical Adjustment image.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...