8 Special procedures
137
Cephalostat CPU FPGA core update
NOTICE! The instructions below are valid when the
cephalostat CPU is currently running a previous version of
the cephalostat software. For instructions on programming
a non-responsive or unprogrammed spare part CPU board,
refer to chapter Special cases.
1.
Press the F7 button and press enter, or type in manu-
ally:
xfr
OP3DPRO
_CEPHR3210_R<ver_major>_<ver_mi-
nor>.srec core
Example:
xfr
OP3DPRO
_CEPHR3210_R2_00.srec core
2.
Wait until the update is completed, power off and power
on again. Do not enter safe mode like in main CPU
R3220 update.
Cephalostat firmware update
NOTICE! The instructions below are valid when the
cephalostat CPU is currently running a previous version of
the cephalostat software. For instructions on programming
a non-responsive or unprogrammed spare part CPU board,
refer to chapter Special cases.
1.
Press the F8 button and press enter, or type in manu-
ally:
xfs
OP3DPRO
_CEPHRFW_R<ver_major>_<ver_mi-
nor>.srec main
Example:
xfs
OP3DPRO
_CEPHFW_R2_00.srec main
2.
Wait for the software package to be transferred and for
the cephalostat CPU to boot itself. If the update fails, try
again.
Verify and set the software version
1.
Wait for the unit to start up and then type:
login
The unit should reply with a list of software versions.
Check that they match the versions listed in the file ver-
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...