7 Cephalometric calibration and alignment
93
7
Cephalometric calibration and
alignment
7.1 Preparations
NOTICE! The nasion support and the ear holders must be
turned out of the way for cephalostat calibrations, so they
are not in front of the x-ray beam.
NOTICE! Check the orientation of the nasion support and
the ear holders so that the patient is always facing to the x-
ray room.
Install the cephalostat unit and make sure that it is in
balance using spirit levels.
NOTICE! Before calibrating the cephalostat update the unit
firmware. See Firmware update instructions.
Take an exposure.
7.2 Cephalometric calibration sequence
1. Balance the cephalostat.
2. Do the primary collimator horizontal adjustment (auto-
matic). Adjust the rotator angle to have the x-ray beam di-
rected exactly towards the cephalometric sensor.
3. Do the rotation angle adjustment (automatic). Adjust the
rotator angle to have the x-ray beam directed exactly to-
wards the cephalometric sensor.
4. Do the mechanical collimator adjustments and check the
cephalostat fixation angle (manual, software guided),
which you can adjust if necessary.
5. Do the secondary collimator horizontal adjustment (auto-
matic). Fine tune the position of the secondary collimator
for optimal alignment with the sensor and X-ray source.
6. Do the cephalostat laser adjustment.
7. Do the cephalometric pixel calibration. Capture sensor
calibration data for performing image corrections.
8. Check the cephalometric quality (optional). Verify that
the image quality is appropriate.
9. Align the ear holders.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...