7 Cephalometric calibration and alignment
113
5. Visually evaluate the result using the installed imaging
software.
Subjects to be evaluated:
1. Smoothness of the exposed area. Non-exposed area
surrounds the whole image.
2. High contrast resolution; minimum 3.1LP/mm must be
distinguishable.
3. All four low contrast holes must be visible.
7.2.9
Ear holder alignment
NOTICE! Rotate the ear holders into lateral view position
and move them completely apart. Push nasion support as
in as it go.
The ear holder center pins and the x-ray focal spot must be
in line.
1. Turn the ear holders into the lateral view position and
move them completely apart.
2. Select the cephalometric exposure program.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...