1 Introduction
7
Always ensure to fulfill the requirements of the local and na-
tional regulations.
The correct software and settings in the workstation are
essential to the performance of the unit. Consult technical
support to ensure correct setup.
Danger - Explosion hazard
Do not use in the presence of flammable anesthetics, gases or
vapors.
The unit is factory set to operate using a 230-240 ±10 VAC
power supply. Never connect the unit to a power supply
different to the voltage marked on the unit.
If the unit needs to be connected to a multiple socket-outlet,
the socket shall not be placed on the floor.
To avoid the risk of electric shock, the unit must only be
connected to a supply mains with protective earth.
The site must fulfill the environmental requirements in the
installation manual chapter technical specifications.
There should be free space around the unit for safe
operation.
The PC / Ethernet switch to which the unit is connected,
should be approved appropriately (e.g. EN 60950, IEC
60950, UL 60950). After installation, check that the IEC
60601-1 leakage current levels are not exceeded.
This product itself complies with IEC 60601-1 medical
safety standard but in order to the system incorporating also
a PC to comply the standard, EITHER the PC has to be a
medical PC OR the PC has to be located over 1,5 meters
apart from the unit. The installer and the user of the system
shall confirm that at least one of the above requirements is
fulfilled. A PC is a medical one if it complies IEC 60601-1
standard and that is indicated in the accompanying
documents of the PC.
The unit shall be connected directly to the acquisition PC
with an Ethernet cable. Connection through the LAN-net-
work of the site is not allowed. Two network ports are
needed in the PC in order to connect also to the site
network.
All service operations must be made by authorized service
personnel only.
The annual service as described in manual is mandatory for
the correct and safe operation of the unit.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...