USER INTERFACE
2.13.2018
97
Command
Parameter
Response of instrument
in_sp_08
kein
Flow rate setpoint.
[not on A30]
in_sp_09
none
Value from pump pressure setpoint.
[not on A30]
in_sp_10
none
Selected variable setting via the serial interface.
in_sp_11
none
Temperature indication in °C or °F.
in_sp_12
none
Pump pressure indication in bar or psi.
in_sp_13
none
Flow indication in l/min or gpm.
in_sp_14
none
Pressure warning limit, upper.
in_sp_15
none
Pressure warning limit, lower.
in_sp_16
none
Pressure alarm limit (5 s).
in_sp_17
none
Pressure alarm limit (1 s).
in_sp_18
none
Flow rate warning limit, upper.
in_sp_19
none
Flow rate warning limit, lower.
in_sp_25
none
Maximum temperature gradient, heating.
in_sp_26
none
Maximum temperature gradient, cooling.
In_hil_00
none
Max. cooling power (%).
In_hil_01
none
Max. heating power (%).
in_mode_01
none
Setpoint for control set to:
0 = Setpoint1
1 = Setpoint2
2 = Setpoint3
In_mode_02
none
Selftuning type:
0 = Selftuning “off”
1 = Selftuning “once”
2 = Selftuning “always”
in_mode_03
none
Type of external programmer input:
0 = Voltage
0 V to 10 V
1 = Current
0 mA to 20 mA
in_mode_04
none
Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
in_mode_05
none
Unit in stop/start condition:
0 = stop
1 = start
in_mode_08
none
Adjusted control dynamics
0 = aperiodic
1 = standard
in_mode_09
none
Selected pump control
0 = Stage control
1 = Pressure control
2 = Flow rate control
3 = Flow rate control (Bypass)