Error messages
114
2.13.2018
Alarm-Code
Cause
Diagnosis / Remedy
E506
Invalid flow value during flow regulation. The calculated flow value is invalid for
the flow regulator, or the function of the
external sensor connection "JULABO
Sensor Pressure/Flow" not set to the
Flow setting.
Please deactivate flow regulation, or
switch the function of the external sensor
connection to Flow.
E507
Invalid setpoint calculation in internal
program sensor.
Please restart the profile.
E508
Invalid setpoint was received on the
internal bus.
Please set a valid setpoint.
E509
Invalid setpoint should be send on the
internal bus.
Please set a valid setpoint.
E510
Current measurement at the "JULABO
Sensor Pres./Flow" sensor input
incorrect.
Check the sensor and the menu settings
for the sensor input "JULABO Sensor
Pressure/Flow".
E513
Flow rate control via external bypass
activated, but external bypass is not
recognized.
Connect an external bypass to the system or
change the type of pump control.