"Connect unit" menu
76
2.13.2018
Important:
The usable temperature range between the >„lower value“ < and
the > upper value < is limited to the configured working
temperature range of the unit (see technical data for working
temperature range).
Select
Push EProg
The current settings are displayed.
0.0 V
equals
50.0 °C
10.0 V equals
250 °C
Push edit
Step 1 is displayed.
Push value
and select.
Push signal type
and select.
Continue automatically to step 2
switch window
Please observe the instructions for
step 2.
Example: 0.0 mA
Set lower value
and
.
Example: -100 °C
Continue automatically to step 3