USER INTERFACE
2.13.2018
53
10.2. „External setpoint“ menu
Select external setpoint
Choose between
>Off<
>Pt100<
>EProg<
REG+E-PROG
In addition to the serial interface via remote control the unit offers
the possibility to adjust the setpoint via analog interface >EXT.
Pt100< or >REG+E-PROG<.
Possible parameters:
Off
- Setpoint is set via the touch screen or via the integrated
programmer. (factory setting)
Pt100 -
Setpoint setting via the analog socket „EXT Pt100“ using
an external temperature sensor or an appropriate
voltage/current source.
EProg
- Can only be adjusted when an electronic module with
analog connections is used (option).
Setpoint setting via the analog interface REG+E-PROG
connection with an external voltage or current source or a
programmer.
Important:
Connect the external voltage or current source or a
programmer to the circulator via the socket REG+E-PROG
(see page 75).