USER INTERFACE
2.13.2018
65
10.5. "Analog interfaces" menu
Select menu
The buttons will display the current
settings.
indicates available submenu
If the electronic module has not been installed, the key
>Analog Module< will not be displayed in this menu.
10.5.1. Alarm output
The buttons will display the current
settings.
Breaking capacity
max. 30 W / 25 VA
with turn-on voltage
max. 30 V
DC
; max. 25 V
AC
with switching current
max. 1 A
Socket on the front
Alarm output: Output for external alarm signal.
This contact is a potential-free change-over contact. All of the unit's
operating conditions can be sent externally via settings in the
>Alarm output< menu without modifying the plug connection.
Setting
Standby
,
Alarm
,
Alarm+Stdby
or
Pump on
connects pins
2 and 3.
Setting
Standby / Inverted
,
Alarm / Inverted
,
Alarm+Stdby /
Inverted
or
Pump on / Inverted
connects pins 2 and 1.