USER INTERFACE
2.13.2018
115
15.2.
Warning messages
Alarm-Code
Cause
Diagnosis / Remedy
E03
Overtemperature limit exceeded during
operation.
The temperature measured is above the
overheating limit. Please raise the
overheating limit, or lower the temperature
setpoint.
E04
Subtemperature limit violated during
operation.
The temperature measured is below the
configured subtemperature limit. Please
lower the subtemperature limit, or increase
the temperature setpoint.
E40
The early warning system for low liquid level
is indicating a critical liquid level.
Please refill temperature control liquid.
E41
The early warning system for high liquid
level is indicating a critical liquid level.
Please drain temperature control liquid.
E1103
Error in level detection.
If this error continues to occur, please
contact JULABO Service.
E1104
Error in level detection.
If this error continues to occur, please
contact JULABO Service.
E1107
The temperature of the heat exchanger or
the internal reservoir is near the configured
safety temperature.
Check the setting of the high temperature
cut-off.
E1108
The temperature of the heat exchanger or
the internal reservoir is near the configured
safety temperature.
Check the setting of the high temperature
cut-off.
E1109
Probe differential limit activated.
Check heat transfer medium for suitability
for use in the temperature range in use!
The viscosity of the heat transfer medium
may not exceed the maximum permitted
value at any working temperature.
Please contact JULABO Service if you have
any questions regarding JULABO heat
transfer media.
E1203
Error in pump pressure measurement.
If this error continues to occur, please
contact JULABO Service.
E1204
Error in pump pressure measurement.
If this error continues to occur, please
contact JULABO Service.
E1205
Error in cooling water flow rate
measurement.
If this error continues to occur, please
contact JULABO Service.
E1206
Error in cooling water flow rate
measurement.
If this error continues to occur, please
contact JULABO Service.
E1207
Error in total current measurement.
If this error continues to occur, please
contact JULABO Service.
E1208
The current measurement for total current is
incorrect.
If this error continues to occur, please
contact JULABO Service.
E1209
Error in current measurement for
compressor stage 1.
If this error continues to occur, please
contact JULABO Service.