USER INTERFACE
2.13.2018
83
11.3. "Adjust sensors" menu
Notice:
Do not alter the factory-setting for the internal sensor!
This is a closed temperature control system: only the calibration of
the external sensor is sensible.
Both the internal temperature sensor and any external temperature
sensor (attached to the "ext. Pt100" socket) may be calibrated.
Principle
: external sensor calibration
During calibration in the external bath, a reference temperature
sensor is used to determine the bath temperature in a stabilized
condition.
This value is then entered under the >Adjust sensors< menu,
>Calibration value< menu item.
1.
Temperature system
2.
External temperature sensor
Pt100
3.
External bath
4.
Temperature measuring
device with a reference
temperature sensor.
(Indicates the calibration
value)
Preparations:
Connect the external sensor Pt100 to the connecting socket „
EXT Pt100 “.
Set unit to >internal control< (see page 23).