"Connect unit" menu
54
2.13.2018
10.3. "Actuating variable" menu
The variable is the degree to which the heater or the refrigeration
unit is activated. The bath is heated or cooled in accordance with
this variable. If this is controlled via the unit's control electronics,
referred to as the >
Controller
<, the bath temperature will be
brought precisely to the selected setpoint and stabilized at that
temperature.
The unit must be in Start mode in order to input variables in the
>
Digital
< and >
EProg
< positions.
Selecting how variables are
inputted.
Choose between inputting
variables via
>
Controller
<
or
>
Digital
<
or
>
EProg
<.
Possible parameters:
Controller
–The internal control electronics of the unit controls the
heater and the connected cooling unit.
Self-tuning is possible. (factory setting)
Digital
–The unit receives the control signal via the digital
interfaces.
Self-tuning is not possible.
EProg -
The unit receives the control signal via the E-Prog input.
Self-tuning is not possible.
- Setting requires electronic module.