USER INTERFACE
2.13.2018
117
Alarm-Code
Cause
Diagnosis / Remedy
E1434
Cooling water temperature too high
Check cooling water.
E1435
Cooling water temperature too low
Check cooling water.
E1501
A timeout occurred on the serial interface.
If the watchdog is activated, the setpoint
must be sent to the device periodically, at
least every 30 seconds.
1502
The pressure in the temperature control
circuit is higher than the configured warning
limit
(1) Correct the pressure setpoint downwards
or
(2) Adjust the warning limit upwards (if
reasonable from a safety standpoint).
1503
The pressure in the temperature control
circuit is lower than the configured warning
limit
(1) Correct the pressure setpoint upwards or
(2) Adjust the warning limit downwards (if
reasonable from a safety standpoint).
1504
The flow in the temperature control circuit is
higher than the configured warning limit
(1) Correct the flow setpoint downwards or
(2) Adjust the warning limit upwards (if
reasonable from a safety standpoint).
1505
The flow in the temperature control circuit is
lower than the configured warning limit
(1) Correct the flow setpoint upwards or
(2) Adjust the warning limit downwards (if
reasonable from a safety standpoint).