USER INTERFACE
2.13.2018
9
ALARM display
Error messages are displayed in a
red box.
Resolution for example E14:
Touch the red box to
mute the alarm.
Press <
> button for help text.
The module and the configuration
are displayed.
Follow the instructions in the help
text.
Press <
> button.
Press <
> button again for help
text.
Press <
> button.
.
Another error message (E108)
appears and describes a pathway
for overcoming the alarm.
Press <
> button for help text.
Follow the instructions in the help
text.
The unit is now ready to continue
operation.
The unit switches to „Standby“. Heater, refrigeration unit and
circulation pump are switched off.
Error message (E 14) remains although the safety temperature
has been raised.
A <Reset> is not permitted in this case because, according to
NAMUR, this condition must be resolved via hardware.
Errors not subject to this regulation are resolved via
<Reset>.
Not all alarms may be removed on-site.
List of all error messages see page 108