"Safety adjustments" menu
50
2.13.2018
9.2.4. Upper and lower pressure warning limit
The buttons will display the current
settings.
Choose button and set value.
Alarm:
Alarm messages are shown in a
red window.
Press <
> button for help text.
For the pressure setpoint following limits must be set.
The >
limit pressure
< sets the upper limit.
Exceeding this pressure for more than 5 seconds, results in an
alarm cut-off and an error message
(Error 302).
Achieving the >
Limit Pressure Peak
< results in an alarm cut-off
and an error message (Error 301).
9.3. Flow limits
The buttons will display the current
settings.
Choose button and set value.
9.3.1. Min. and max. flow setpoint
The buttons will display the current
settings.
Choose button and set value.
Min. and max. setpoint:
Limits the selectable flow range.
The selected flow values must be between the limit values defined
here.