"Install unit" menu
86
2.13.2018
Examples:
1-point< calibration
3-point< calibration
°C
°C
T 1
T
°C
°C
T 1
T
T
T
1 = Original curve
In case of a 1-point-calibration the calibration curve is moved
entirely towards the original curve of the sensor.
In case if a 3-point calibration a bent curve may result. Thus the
accuracy of the temperature indication can be improved in areas
important to the application.
Example:
Working temperature setpoint 150 °C
The comparison points can be set at 140 °C, 150 °C and 160 °C.
11.4. "Reset unit" menu
and