"Connect unit" menu
72
2.13.2018
REG+E-PROG connector
REG+E-PROG
Only SELV entrance
(Safety Extra Low Voltage)
Three logging outputs and one input for an external programmer:
1 Chan. 1 voltage output
0..10 V
2 Chan. 2 voltage output
0…10 V
3 Gnd
for outputs
0 V
4 EProg external programmer input
0…10 V / 0…20 mA
5 Chan. 3 current output
4…20 mA / 0…20 mA
6 Gnd
for programmer
0 V
Outputs of the connector
(Description on page 74)
(voltage output)
Push value
First define output value for
channel 1.
Example:
Confirm with
Define scale:
Assign the value to 0 V.
and
Assign the value to 10 V
and
Assign the lowest value which is
to be emitted to 0 V,
the highest to 10 V (in the
example on the right: °C).
The setting is displayed on the
keypad.
Examples: