Communication via Modbus TCP/IP
104
2.13.2018
Register-
address
Protocol
Address
Datatype Description
Adjustable range
16
ushort
control dynamics
0: Aperiodic
1: Standard
17 – 18
float
Xp control parameter of the
cascade controller
0.1 … 99.9 1/K
19
ushort
Tn control parameter of the
cascade controller
3 … 9999 s
20
ushort
Tv control parameter of the
cascade controller
0 …999 s
21 – 22
float
XpU control parameter of the
cascade controller
0.1 … 99.9 1/K
23 – 24
float
CoSpeed for external control
0.0 … 5.0
Controller limits
40
short
maximum cooling power
-100 … 0
41
short
maximum heating power
0 … 100
42 – 43
float
Min. internal temperature of the
cascade controller
44 – 45
float
Max. internal temperature of the
cascade controller.
46
short
Lower band limit
0 … 200 K
47
short
Upper band limit
0 … 200 K
Pump settings
50
ushort
control type pressure control /
stage control
0: Stage control
1: Pressure control
2: Flow control
3: Flow control (Bypass)
51
ushort
Pressure control of
internal/external pressure sensor
0: Pressure control of
internal sensor
1: Pressure control of
external sensor
52
ushort
Pumpstage
1 – Pumpstage max. (74)
53 – 54
float
Working pressure
(Pressure setpoint)
0 … Pressure setpoint max
(75)
55 – 56
float
Working flow rate
(Flow rate setpoint)
Temperature limits
60 – 61
float
Low temperature warning limit
(SubTemp)
62 – 63
float
High temperature warning limit
(Overtemp)
64 – 65
float
float maximum temperature
gradient, cooling
-100 … -0.1
- infinite (0xFF800000) for
deactivation
66 – 67
float
maximum temperature gradient,
heating
0.1 … 100
+ infinite (0x7F800000) for
deactivation
3
The dynamics cannot be changed with the maximum temperature gradient, heating/cooling active.
4
If the maximum temperature gradient, heating/cooling, values change, the dynamics of the internal control is
automatically adjusted to “Standard”