P/N 960-100929R_Rev. 3 {EDP #148849}
© 2012, Japan CashMachine Co., Limited
6 - 1 0
Section 6
iVIZION® SeriesNext-Generation Banknote Acceptor Unit
Calibration and Testing
20. When the Calibration Value writing EEPROM
writing is complete, the “
Calibration Informa-
tion
” Screen shown in Figure 6-50 will re-appear.
21. Mouse-click on the “
OK
” Screen Button
(See Figure 6-50 a).
This completes the Paper Calibration Procedures.
Refer to “Reference Paper Use Precautions” on
page 1-3 of Section 1 to properly store and prevent
damage to the KS-072 Reference Paper.
C
ALIBRATION
P
LUS
S
ERIAL
N
UMBER
W
RITING
When the Sensor Calibration procedure is com-
plete, the capability to write the Serial Number into
the iVIZION
®
Unit becomes possible. To write a
Serial Number into Memory, proceed as follows:
1. Complete the initial Sensor Calibration Proce-
dure first (See “Calibration Only” on page 6-7 of
this Section).
2. Once the CIS Sensor Calibration is complete, the
“
iVIZION Calibration Service Suite Edition
”
Screen shown in Figure 6-51 will appear.
3. Mouse-click on the “
OK
” Screen Button
(See Figure 6-51 a). The “
Setting Manufacture
No
” Screen shown in Figure 6-52 will then
appear.
4. Mouse-click on either the “
” or “
” Triangle
Indication Buttons in each Serial Number Box
located at the bottom of the Screen to either
increase or decrease the numbers, or change the
existing numbers using the “
” or “
” Triangle
Button (See Figure 6-52 a).
5. Mouse-click on the “
OK
” Screen Button
located on the right side of the Screen
The Calibration Value and the Serial Number will
then be written into the EEPROMs on the Sensor
and the CPU Boards (See Figure 6-53).
When the data writing into the EEPROM is
complete, the “
Calibration Information
” Screen
shown in Figure 6-54 will re-appear.
6. Mouse-click on the “
OK
” Screen Button
This completes the Serial Number Writing
Procedures.
Performance Tests
This portion explains the iVIZION
®
Performance
Test Procedures. The following two (2) methods
exist to perform this Performance Test Procedure.
•
Performance Test using a PC
•
Performance Test using the External DIP Switches.
Choose one (1) of the two (2) above Performance
Test Procedures by selecting the one related to the
particular circumstance desired.
Figure 6-49
Calibration Value Writing Screen
NOTE: The Barograph does not indicate a Calibration in progress.
Figure 6-50
Calibration Complete Screen
a
NOTE: To perform the “Writing Serial No.”,
process, Mouse-Click on the “
Update
serial No.
” Check-box when the “iVIZION
Calibration Ver.X.XX” Screen is active
(See iVIZION Calibration Ver.X.XX
Screen Figure 6-36 b on page 6-8 of this
Section).
Figure 6-51
iVIZION Calibration Service Suite
Edition Screen
a
Figure 6-52
Setting Manufacture No. Screen
a
b
Figure 6-53
Calibration Value Writing Screen
NOTE: The Barograph does not indicate a Calibration in progress.
Figure 6-54
Adjustment Info Screen 8
a