6
Lift the damaged selector knob from the
membrane panel assembly.
7
Push the replacement knob on the
membrane panel switch - align with flat on
switch .
8
Install the new bezel by tightening the
included (7) 4-40 X ¼ screws.
9
Place the front case assembly over the vent
module and tighten the (2) 4-40 X ¼ screws
unto the dovetail mounting bracket.
Reconnect the ribbon cable unto the PIM
PCB. Make sure the two locking “ears” lock
into position.
10
Attach the back case to the vent module
and align cover with handle, air intake
housing and dovetail mounting studs.
Insert and tighten the included (2) 10-32
Keps nuts and the (4) 6-32 X 2” screws.
11
Tighten the (4) 8-32 X 3 screws on the
outer air intake.
12
For Eagle II ventilators the Selector knob
can be removed by prying the knob up
from the case. Align flats on knob and shaft
and push down to install.
13
Perform HiPot Testing then perform
Functional Test using the RCS.