906-0731-04 Rev. B Sept. 2012
Page 65 of 68
2100
Patient Disconnect
Paw < [PEEP + 7 cmH2O]
3110
RTC Battery Fault (RTC Battery Low)
Vstby < 2.5V
712-0731-02
3120
Self Check Fault (Calibration Due)
> 365 days since last calibration
Recalibrate
using RCS
3130
Ambient Pressure Fault (Excessive
Altitude Sensor Failure)
ambient pressure transducer
has failed
712-0731-02
3131
Ambient Pressure Fault (Excessive
Altitude)
ambient pressure transducer:
altitude > 25,000 ft
Ventilator
being used
outside of
specification
limits
3132
Ambient Pressure Fault (Excessive
Altitude)
ambient pressure transducer:
altitude < -2,000 ft
3140
Operational Temperature Fault
ambient temperature > 55C
3141
Operational Temperature Fault
(Excessive Temperature Low)
ambient temperature < -10C
3143
Self Check Fault
failure of internal temperature
sensors
712-0731-02
2170
Spontaneous Breath - PIP High
Paw > high limit during 2
consecutive spontaneous
breaths
2171
Spontaneous Breath - PIP Low
Paw < low limit during 2
consecutive spontaneous
breaths
1172
Run-Time Self Check Alarm
supply voltage to circuitry NG
712-0731-02
2172
Spontaneous Breath - Vt High
Vt > high limit during 2
consecutive spontaneous
breaths
1173
Internal Comm Failure
communication between EMV
and SPM fails
Check Cable
from PIM
board to CPU;
702-0731-02
2173
Spontaneous Breath - Vt Low
Vt < low limit during 2
consecutive spontaneous
breaths
1174
Self Check Failure
failure of PGA offset control
detected during startup
712-0731-02
1175
Internal Comm Failure
SPM DSP unable to
communicate with PGA's
712-0731-02
1176
Offset Self-check Failure
calibration file fails its integrity
test
712-0731-02
2300
Pulse Ox Module Failed
712-0731-14
3300
Pulse Ox Module Failed
2301
Internal Comm Failed
communication between pulse
oximeter module and CPU fails
712-0731-14
3301
Internal Comm Failed (Comm Failure
EMV-Pulse Ox - Monitor Not In Use)
communication between pulse
oximeter module and CPU fails
3310
Pulse Ox Sensor Not Connected
pulse oximeter detects SpO2
sensor disconnect; no power
from either internal battery or
external source
712-0731-14
3311
Defective Pulse Ox Sensor (Defective
Sensor)
3312
Pulse Search (Pulse Search)
3313
Pulse Ox Signal Interference