906-0731-04 Rev. B Sept. 2012
Page 42 of 68
5
Remove the O2 Inlet fitting by
unscrewing the dust cap then unscrewing
the (3) 8-32 X 7/16 screws.
6
Loosen and remove the (3) 8-32 X ¼
screws. Loosen and remove the Gas
Output fitting using a 1-inch deep-socket
wrench.
7
Disconnect the Power Input cable by
pressing on the locking latch and pulling
the cable straight up from the connector.
8
Loosen and remove the (2) 4-40 X 1 ¼
Screws and nylon spacers supporting the
Power Input assembly unto the chassis.
9
Loosen and remove the (2) 6-32 Keps
nuts using a 5/16 open-end wrench.
10
Using a needle nose pliers, carefully
remove the 3 tubing on the
“Transducer”, “Exhaust Do not Occlude”,
and “Exhalation Valve” fittings.
11
Loosen and remove the (2) 6-32 Keps
nuts using a 5/16 open-end wrench.