906-0731-04 Rev. B Sept. 2012
Page 55 of 68
Power Input Kit 712-0731-17
Contents:
Instructions:
Step No.
Directions
1
Loosen but do not remove the
(4) 8-32 X 3 screws on the outer
air intake.
2
Loosen and remove the (2) 10-32
Keps nuts and the (4) 6-32 X 2
screws. Remove the back case by
lifting from the ventilator.
3
Remove the (2) 4-40 X ¼ screws
on the Dovetail Mounting
Bracket and disconnect the
ribbon cable on the PIM PCB by
simultaneously applying pressure
on the two locking “ears”.
4
Flip the ventilator over and
remove the front case assembly
by lifting it straight up away from
the ventilator module.
ITEM
QTY
DESCRIPTION
1
Power Input Assembly
2
Screw, Phillips, Flat Head, Zinc Plated, 4-40 X 1/4
2
Screw, Phillips, Flat Head, 4-40 X 1 1/4
2
Spacer, Nylon, #4, x .812 Long