Gas Output Kit 712-0731-16
Contents:
Instructions:
Step No.
Directions
1
Remove the damaged fitting by unscrewing
with a 1-inch deep socket wrench.
2
Place O-ring unto underside of outlet fitting.
3
Place fitting and O-ring unto Gas Output and
tighten with a 1” deep socket wrench. (Do
not cross thread).
4
Perform HiPot Testing then perform
Functional Test using the RCS.
ITEM
QTY
DESCRIPTION
1
Fitting, Patient, Outlet, SPM
1
O-Ring, Neoprene, 1/2" ID