906-0731-04 Rev. B Sept. 2012
Page 18 of 68
4
Flip the ventilator over and remove
the front case assembly by lifting it
straight up away from the ventilator
module.
5
Remove the bezel by loosening and
removing the (7) 4-40 X ¼ screws
that hold the bezel to the front case.
6
Using a sharp knife, carefully cut the
RTV sealant around the USB Printed
Circuit Board and around the SPO2
Connector. Be careful not to cut
either cable.
7
Loosen and remove the two 4-40 X
3/16 screws holding the Mini USB
Cable Assembly to the front case.
Loosen and remove the two M2.5 X
5mm screws holding the SPO2 cable
to the front case.
8
Loosen and remove the two (2)
6-32 X 5/16 screws that hold the
CPU/UIM & SPO2 Stack to the front
case.
9
Lift the CPU/UIM & SPO2 Stack up
from the front case. Handle the
SPO2 cable with extreme care. Do
not pull on the cable.
10
Loosen and remove the two 4-40 X
5/16 screws holding the USB
Connector Plate to the front case.