i-SmartCare 10 Blood Gas Analyzer
QC Sample
,
continued
4. Sample Analysis
Introduce
QC Sample,
continued
4. “
Aspirating QC sample. Please wait…
” message appears as the QC
Sample is being processed.
5. Once the aspiration is complete, “
Remove QC sample now
” message
will appear. Please remove the QC sample from the sampler probe at
once.
6. Please wait until the sampler is automatically closed.
Once the aspiration is successfully finished, the screen will change to
QC-Information
.
Note:
If the QC sample aspiration process is not done properly or the QC
sample aspiration amount is insufficient, the message “
Insufficient
QC sample error. QC sample is being aborted
.” appears.
7. The QC - Information screen will appear.
Note:
Items with units will display with the units selected from
Setup -
Unit
.
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...