7. Maintenance
114
i-SmartCare 10 Blood Gas Analyzer
Cleaning
Caution
•
To prevent infection, please wear protective clothes and gloves while
cleaning.
•
Perform routine cleaning of the analyzer.
•
Use 0.5 % sodium hypochlorite solution to clean the analyzer
•
Do not spray cleaning solution directly on the analyzer.
•
When using cleaning solution, make sure it does not get inside ofthe
analyzer.
•
Do not apply too much force when wiping the analyzer’s screen
•
Dispose any biohazard materials after cleaning in accordance with the
lab’s guidelines and procedure.
Cleaning procedure
1. Use 0.5% sodium hypochlorite solution.
Note:
Commercial bleach contains approximately 5% sodium
hypochlorite.
Add water in 1:9 ratio to dilute the solution before use.
2. Soak a soft cloth in the cleaning solution.
3. Clean the sampler cover, screen, and other contaminated parts of the
analyzer with the cloth.
4. Wait for 10 minutes.
5. Wipe all surfaces with a wet cloth (water).
6. Finally, wipe the unit with a dry cloth.
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...