Operator's Manual
125
Operation Specifications
,
continued
9. Product Specifications
Input Parameters
•
Input Parameters and Ranges
Parameter
Description
Unit
Input range
Default
Patient Temp.
Patient
temperature
°C
15.0 ~ 45.0
37.0
tHb
Total
hemoglobin
g/dL
0.0 ~ 28.0
-
s
O
2
Oxygen
saturation
%
0 ~ 100
-
F
IO
2
Fraction
of inspired
oxygen
%
0 ~ 100
21
BP
Barometric
pressure
mmHg
540 ~ 800
760
Calculated
Parameters
•
Calculated Parameters and Ranges
Parameter
Description
Unit
Input range
pH(
T
)
Temperature corrected pH
-
6.000 ~ 8.500
p
CO
2
(
T
)
Temperature corrected
p
CO
2
mmHg
0.0 ~ 250.0
HCO
3
-
Bicarbonate
mmol/L
0.0 ~ 99.0
HCO
3
-
(std)
Standard bicarbonate
mmol/L
0.0 ~ 99.0
BE(B)
Base excess of blood
mmol/L
-50.0 ~ 50.0
BE(ecf)
Base excess of extracellular
fluid
mmol/L
-50.0 ~ 50.0
tCO
2
Total CO
2
mmol/L
0.0 ~ 99.0
p
O
2
(
T
)
Temperature corrected
p
O
2
mmHg
0 ~ 800
p
O
2
(A-a)
Alveolar-arterial
p
O
2
gradient
mmHg
0 ~ 700
tHb
Total hemoglobin
g/dL
0.0 ~ 28.0
s
O
2
Oxygen Saturation
%
0 ~ 100
Anion gap
Anion Gap
mmol/L
0 ~ 75
Ca
2+
(7.4)
Ionized Calcium adjusted to
a pH of 7.4
mmol/L
0.22 ~ 5.58
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...