i-SmartCare 10 Blood Gas Analyzer
Troubleshooting
,
continued
8. Troubleshooting
Printer
In case you experience one of the following issues:
•
The printer does not print.
•
The printer does not feed paper.
1. Check if the connection between the analyzer and the power outlet is
loose.
2. Open the printer cover.
3. If the printer is out of paper, load paper.
4. If the paper is jammed, remove the jammed paper and press the
RESET
switch located on the left side of the printer.
5. Connect the power adapter.
When the analyzer is running on a battery and the battery level is below
80 %
, the print function is not available.