i-SmartCare 10 Blood Gas Analyzer
Blood Sample
,
continued
Introduce Sample,
continued
4. Sample Analysis
3. In
Capillary
mode, carefully place the capillary tube in the septum and
touching the screen to aspirate sample.
Note:
Make sure there is no gap between the capillary tube and the
septum so that no air bubbles are introduced to the sampler.
4. “
Aspirating sample. Please wait…
” message appears as the sample is
aspirated into the analyzer.
5. When the sample aspiration is complete, the message “
Remove
sample now
” appears. Please remove the sample container from the
sampler probe at once.
6. Please wait until the sampler is automatically closed.
Once the aspiration is successfully completed, the screen changes to
Sample – Information
.
Note:
If the sample aspiration process is not done properly or the sample
aspiration amount is insufficient, the message “
Insufficient sample
error. Sample is being aborted.
” appears.
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...