2. Installation
18
i-SmartCare 10 Blood Gas Analyzer
Power-On
Caution
•
Before turning on the analyzer, please remove all USB memory or
network cable from the analyzer.
Turn on the power
1. Turn on the power by pushing the switch on the back.
2. The LCD screen is turned on and the analyzer will start booting.
3. Once the booting is complete, the main screen appears.
Note:
When the analyzer is started without a cartridge, “No cartridge”
status appears.
If the unit is rebooted with a usable cartridge, “Cartridge Restart”
automatically performs.
If the unit is rebooted with a unusable cartridge, “Expired Cartridge”
status appears.
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...