Operator's Manual
57
Sample Setup
,
continued
3. Instrument Settings
Correlation Factors
Setup,
continued
Note:
Acceptable ranges for slopes and offsets of the correlation factors
are as follows:
Parameter
Slope range
Offset range
pH
0.00 ~ 9.99
±99.9999
p
CO
2
0.00 ~ 9.99
±99.99
p
O
2
0.00 ~ 9.99
±99.9
Na
+
0.00 ~ 9.99
±99.9
K
+
0.00 ~ 9.99
±99.99
Ca
2+
0.00 ~ 9.99
±99.999
Cl
-
0.00 ~ 9.99
±99.9
Hct
0.00 ~ 9.99
±99.9
Glu
0.00 ~ 9.99
±99.9
Lac
0.00 ~ 9.99
±99.99
4. Press for further options.
•
Press
button to print out the set correlation factors.
•
Press
button will reset all the entered values to default ones.
5. Press
to save the settings and to exit to the main screen.
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...