108
i-SmartCare 10 Blood Gas Analyzer
Cartridge Removal
,
continued
7. Maintenance
Remove cartridge
in use
1. To remove a non-expired cartridge, select
MENU > CARTRIDGE >
Replacement
and the following message appears.
Note:
To cancel cartridge removal, press
NO
.
Once the cartridge is removed, it cannot be reused even if the
expiration date and the number of tests are still valid.
When the analyzer is running on a battery and the battery level is
below
40 %
, the cartridge replace is not available.
2. Press
YES
, and the following screen will display.
3. Start the cartridge removal procedure following the instructions on the
screen.
Note:
Do not put the cartridge cover back on the removed.
The cartridge solution may leak out.
4. To install a new cartridge, refer to
2. Installation > Cartridge
Installation
for detailed procedure.
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...