9. Product Specifications
124
i-SmartCare 10 Blood Gas Analyzer
Operation Specifications
Measuring
Parameters
•
Measured Parameters and Ranges
Parameter
Unit
Reportable Range
Resolution
pH
-
6.500 ~ 7.800
0.001
p
CO
2
mmHg
5.0 ~ 150.0
0.1
p
O
2
mmHg
10 ~ 680
1
Na
+
mmol/L
80 ~ 200
1
K
+
mmol/L
1.0 ~ 20.0
0.1
Ca
2+
mmol/L
0.25 ~ 5.00
0.01
Cl
-
mmol/L
50 ~ 150
1
Hct
%
10 ~ 70
1
Glu
mg/dL
5 ~ 500
1
Lac
mmol/L
0.2 ~ 15.0
0.1
•
Sample Type : Whole blood
•
Anticoagulant : Heparin
•
Sample volume : 100 μL
•
Sample introduction method : Aspiration
•
Sample analysis time : 50 seconds
•
Sample analysis temperature : 37.0 ± 0.2 °C
•
Measuring principle : electrochemical
(potentiometry, amperometry, conductometry)
•
Calibration : Automatic or manual
Environmental
requirements
•
Operating location : Indoor, flat surface
•
Operating temperature : 15 ~ 35 °C
※
It has been confirmed that i-SmartCare 10 is safe to operate in the
range of 5 ~ 40 °C according to IEC 61010-1. The recommended range
is from 15 to 35 °C for the desirable performance.
•
Operating humidity : 5 ~ 85 % (relative humidity)
•
Operating altitude : Below 3,000 m
•
Power supply : 100 ~ 240 Va.c.(grounded electrical supply)
50-60 Hz, 1.5 A
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...