6. Shutdown
102
i-SmartCare 10 Blood Gas Analyzer
Power Recovery
Cartridge Restart
•
Even if the power is off during the usage, cartridge can be re-started upon
rebooting the analyzer with the following conditions:
The analyzer is in Ready state or in process of calibration when the
power is interrupted, and the power is restored within 12 hours.
The analyzer is analyzing a sample when the power was interrupted
and the power is restored within 20 minutes.
The analyzer was measuring the QC when the power was out, and
the power was restored within 12 hours.
•
There is no restart limitation of restarts from one cartridge.
•
If the analyzer is turned off longer than 12 hours, the cartridge will
expire.
1.
Cartridge Restart
screen appears once the Cartridge restart is in
progress.
2.
Cartridge Restart
normally takes about 19 minutes.
3. Once completed, the analyzer returns to the
Ready
status.
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...