Operator's Manual
119
Troubleshooting
,
continued
8. Troubleshooting
Power
•
Power connected icon
is not indicated while power cable is
plugged into the analyzer.
1. Check if the connection between the analyzer and the power cable is
loose.
2. Fix the loose power connection.
In case you experience one of the following issues:
•
The analyzer is turned off.
•
The analyzer is not turning on.
1. If the analyzer’s power switch is on, turn it off.
2. Check if the connection between the analyzer and the power outlet is
loose. Fix the loose power connection.
3. Turn on the analyzer.
4. In case the analyzer is not turning on, turn off the power switch again.
5. Charge the battery for approximately 10 minutes.
6. Turn on the power switch of the analyzer again.
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...