8. Troubleshooting
116
i-SmartCare 10 Blood Gas Analyzer
Troubleshooting
Guidelines
•
If the problems described below occur during installation and/or
operation of the analyzer, refer to the solutions as described in this
manual.
•
Contact the reseller or service engineer if the recommended solutions
do not solve the problems.
Barcode Scanner
•
There is no red light for the barcode scanner.
1. Please place a barcode under the part where
is labeled.
2. If you still do not see the red light, insert a sharp object (e.g. a
paperclip or a mechanical pencil tip) into the SCAN hole next to the
DISK LED on the back of the analyzer and press once.
3. The scanner may not automatically recognize a barcode in the dark.
4. If the problem persists, please turn off and restart the analyzer.
Note:
Direct eye contact with the infrared light may cause retinal injury.
Do not look directly into the red light.
Screen
In case you experience one of the following issues:
•
The screen does not respond to touching
•
The screen is frozen
•
Abnormal screen is conditions are detected
1. Turn off the power.
2. Wait for 10 seconds.
3. Turn on the power back.
Summary of Contents for i-SmartCare 10
Page 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Page 3: ......
Page 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Page 5: ......
Page 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Page 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Page 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Page 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Page 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Page 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Page 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...