Horizontal System
Sweep speeds
5 s/div to 2 ns/div main and delayed (HP 54600B, HP 54601B,
HP
54602B
5 s/div to 5 ns/div main and delayed (HP 54603B)
Accuracy
±
0.01%
±
0.2% of full scale
±
200 ps
Vernier
Accuracy
±
0.05%
Horizontal resolution
100 ps
Cursor accuracy
1,2
(
∆
t and 1/
∆
t)
±
0.01%
±
0.2% of full scale
±
200 ps
Delay jitter
10 ppm
Pretrigger delay (negative time)
≥
10 divisions
Posttrigger delay (from trigger point to start of sweep)
at least 2560
divisions or 50 ms. Not to exceed 100 s.
Delayed sweep operation
Main sweep Delayed sweep
5 s/div to 10 ms/div up to 200 times main sweep
5 ms/div and faster up to 2 ns/div
1
Use full scale of 50 ns on 2 ns/div range.
2
Tested, see "To verify horizontal
∆
t and 1/
∆
t accuracy," on page 4–16.
Performance Characteristics
Horizontal System
5–5
Summary of Contents for HP 54600B
Page 7: ...vi ...
Page 12: ...1 The Oscilloscope at a Glance ...
Page 28: ...2 Operating Your Oscilloscope ...
Page 69: ...2 42 ...
Page 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Page 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Page 153: ...4 64 ...
Page 154: ...5 Performance Characteristics ...
Page 175: ...Glossary 8 ...