To perform the self-calibration
In this procedure you load the default calibration factors to give a known
starting point for the firmware calibration.
However, once the default
calibration factors are loaded, you must perform the remainder of the
firmware calibration to maintain the accuracy of the oscilloscope.
Table 4-14
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Pulse generator
100 kHz, 1 V p-p, rise time <5 ns
HP 8112A
Cable
BNC, 3 feet
HP 10503A
Cable
BNC, 9 inches, Qty 2
HP 10502A
Adapter
BNC tee (m) (f) (f)
HP 1250-0781
Adapter
BNC (f-f)
HP 1250-0080
Feedthrough
50
Ω
, BNC (m) and (f)
HP 10100C
1
Check the rear panel
DC CALIBRATOR
output level.
If you are not sure how to check the
DC CALIBRATOR
, see "To check the
output of the
DC CALIBRATOR
," on page 4–6.
2
Load the default calibration factors.
a
Set the rear-panel
CALIBRATION
switch to
UNPROTECTED
(up position).
b
Press
Print/Utility
, then press the
Self Cal Menu
softkey.
c
Press the
Load Defaults
softkey.
Self-calibration hint
Thie instrument is self-calibrated at the factory. However, it should be
self-calibrated again in its working environment after a 30-minute warmup to
obtain the best accuracy.
Service
Adjusting the Oscilloscope
4–31
Summary of Contents for HP 54600B
Page 7: ...vi ...
Page 12: ...1 The Oscilloscope at a Glance ...
Page 28: ...2 Operating Your Oscilloscope ...
Page 69: ...2 42 ...
Page 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Page 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Page 153: ...4 64 ...
Page 154: ...5 Performance Characteristics ...
Page 175: ...Glossary 8 ...