To verify voltage measurement accuracy
In this test you verify the voltage measurement accuracy by measuring the
output of a power supply using dual cursors on the oscilloscope, and
comparing the results with a multimeter.
Test limits
±
1.9% of full scale (HP 54600B, HP 54601B, HP 54602B)
±
2.4% of full scale (HP 54603B)
Table 4-3
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Power supply
14 mV to 35 Vdc, 0.1 mV resolution
HP 6114A
Digital multimeter
Better than 0.1% accuracy
HP 34401A
Cable
BNC, Qty 2
HP 10503A
Shorting cap
BNC
HP 1250-0774
Adapter
BNC (f) to banana (m)
HP 1251-2277
Adapter
BNC tee (m) (f) (f)
HP 1250-0781
Service
Verifying Oscilloscope Performance
4–7