To adjust the low-frequency compensation
In this procedure you adjust the low-frequency compensation adjustment for
each channel.
Table 4-15
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Square wave generator
30 kHz at about 3 Vp-p
HP 8112A
Feedthrough
50
Ω
, BNC (m) and (f)
HP10100C
Cable
BNC
HP 10503A
1
Turn off the oscilloscope and disconnect the power cord.
2
Remove the cover from the oscilloscope.
3
Reconnect the power cord and turn on the oscilloscope.
4
Using the
BNC
cable and 50-
Ω
feedthrough, connect the square wave
generator to channel 1.
5
Adjust the generator for about 30 kHz at about 3 Vp-p.
6
Press
Autoscale
.
7
Set channel 1 to 500 mV/div.
You must perform this adjustment on the 500 mV range.
8
Adjust the output of the generator until you obtain about 5 to 6
divisions of vertical deflection.
Service
Adjusting the Oscilloscope
4–33
Summary of Contents for HP 54600B