9
Adjust the channel 1 low-frequency compensation adjustment for as
flat a pulse top as possible.
10
Repeat steps 4 through 9 for channel 2 (channels 2 to 4 on the
HP 54601B and HP 54602B).
Low-frequency compensation adjustment locations
Low-frequency pulse
response adjustments
Figure 4–3
Service
Adjusting the Oscilloscope
4–34