To troubleshoot Option 005
(HP 54602B only)
To isolate a malfunction to the Option 005 board, do the following:
1
Disconnect the three cables that connect the Option 005 board to the
system board.
2
Verify proper oscilloscope operation, as described in this chapter.
3
If the oscilloscope passes the performance verification and the
malfunction still occurs when the Option 005 board is reconnected,
then you should replace the Option 005 board.
Service
Troubleshooting the Oscilloscope